Specifications |
|
Infusion Rate |
ml/h mode: 1~1200ml/h (1ml/h step) drop/min mode: 1~266drop/min (1drop/min step) |
Purge Rate |
100ml/h---1200ml/h (can be calibrated) |
Flow Rate Accuracy |
ml/h mode: Within ±5% (after correct calibration) drop/min mode: Within ±3% (after correct calibration) |
Mechanical Precision |
Within ±2% |
Maximum infusion pressure |
150kPa |
KVO Speed |
1ml/h---5ml/h (1ml/h step) |
Volume Limit |
1ml---9999ml |
Total Infusion Volume |
0---9999.9ml |
Alarm Pressure
|
High 800mmHg±200mmHg (106.7kPa±26.7kPa) |
Medium 500mmHg±100mmHg (66.7kPa±13.3kPa) |
|
Low 300mmHg±100mmHg (40.7kPa±13.3kPa) |
|
Audible and Visible Alarms |
There are bubbles, tube blocked, door closed, IV set Mismatched, battery running out, infusion finish, equipment failure , abnormally operation, software exception, low battery, forget operations, etc. |
Bubble Detector |
Ultrasonic wave detector; detection sensitivity ≥25μL |
Forbidden |
Not support blood Infusion |
Transport & Storage Condition |
Temperature: -20ºC~+55ºC |
Relative Humidity: ≤95% |
|
Atmospheric Pressure: 50.10KPa---106.0KPa |